Implementation of dosimetry equipment and phantoms at the MedAustron light ion beam therapy facility

2017

Purpose: To describe the implementation of dosimetry equipment and phantoms into clinical practice of light ion beam therapy facilities. This work covers standard dosimetry equipment such as computerized water scanners, films, 2D-array, thimble and plane parallel ionization chambers, but also dosimetry equipment specifically devoted to the pencil beam scanning delivery technique such as water columns, scintillating screens or multi-layer ionization chambers. Method: Advanced acceptance testing procedures developed at MedAustron and complementary to the standard acceptance procedures proposed by the manufacturer are presented. Detailed commissioning plans have been implemented for each piece…

medicine.medical_specialtyIon beamComputer scienceNuclear engineeringmedicine.medical_treatmentImaging phantomPelvis030218 nuclear medicine & medical imagingMedical physicist03 medical and health sciences0302 clinical medicinemedicineHumansDosimetryMedical physicsRadiometryPencil-beam scanningIonsPhantoms ImagingRadiation DosimetersUncertaintyGeneral MedicinePencil (optics)Radiation therapy030220 oncology & carcinogenesisSoftwareMedical Physics