0000000000042420

AUTHOR

S. Vatnitsky

Implementation of dosimetry equipment and phantoms at the MedAustron light ion beam therapy facility

Purpose: To describe the implementation of dosimetry equipment and phantoms into clinical practice of light ion beam therapy facilities. This work covers standard dosimetry equipment such as computerized water scanners, films, 2D-array, thimble and plane parallel ionization chambers, but also dosimetry equipment specifically devoted to the pencil beam scanning delivery technique such as water columns, scintillating screens or multi-layer ionization chambers. Method: Advanced acceptance testing procedures developed at MedAustron and complementary to the standard acceptance procedures proposed by the manufacturer are presented. Detailed commissioning plans have been implemented for each piece…

research product

Patient-specific plan QA in active pencil beam scanning with proton and carbon ion beams: A new software solution to support treatment planning workflow

research product

End-to-end tests using alanine dosimetry in scanned proton beams

This paper describes end-to-end test procedures as the last fundamental step of medical commissioning before starting clinical operation of the MedAustron synchrotron-based pencil beam scanning (PBS) therapy facility with protons. One in-house homogeneous phantom and two anthropomorphic heterogeneous (head and pelvis) phantoms were used for end-to-end tests at MedAustron. The phantoms were equipped with alanine detectors, radiochromic films and ionization chambers. The correction for the 'quenching' effect of alanine pellets was implemented in the Monte Carlo platform of the evaluation version of RayStation TPS. During the end-to-end tests, the phantoms were transferred through the workflow…

research product

Characterization of PTW-31015 PinPoint ionization chambers in photon and proton beams

The increased use of complex forms of radiotherapy using small-field photon and proton beams has invoked a growing interest in the use of micro-ionization chambers. In this study, 48 PTW-TM31015 PinPoint-type micro-ionization chambers that are used in the commissioning and patient specific QA of a proton pencil beam scanning (PBS) delivery system have been characterized in proton and high-energy photon beams. In both beam modalities, the entire set of PinPoint chambers was characterized by imaging them, by evaluating their stability using check source measurements, by experimentally determining the ion recombination, polarity effect and by cross calibrating them in terms of absorbed dose to…

research product