Importance of calibration for diameter and area determination by intravascular ultrasound
Intravascular ultrasound (IVUS) permits quantitative assessment of the lumen diameter and area of coronary arteries. The experimental study was performed to evaluate the accuracy of diameter and area measurements.Lumen quantitation (lumen diameter D and cross-sectional area A) in lucite tubes (lumen diameter 2.5 to 5.7 mm, Plexiglas) was performed using a mechanical IVUS system (HP console, 3.5 F catheter, Boston Scientific, 30 MHz). The influence of fluid type (blood, water and saline solution), fluid temperature (20 degrees C/37 degrees C), catheter to catheter variation, gain setting and ultrasound frequency (12, 20 and 30 MHz) was determined. In blood at 20 degrees C there was a constan…