Acute performance evaluation of a new ventricular automatic capture algorithm
Aims This study evaluated the acute clinical performance of a new ventricular automatic capture algorithm developed to work with all lead types and pacing vectors. Methods and results During regular pacemaker implant or replacement, AutoThreshold and manual threshold tests were performed in ventricular unipolar (UP) and bipolar (BP, if applicable) pacing using a customized external prototype INSIGNIA™ pacemaker. The success rate and accuracy of two different modes (commanded and ambulatory) of the automatic capture algorithm were used to evaluate the performance. Loss-of-capture events (two consecutive non-captured beats without backup pacing) were used to assess safety. Data of 53 patients…
Evolution Over Time of Ventilatory Management and Outcome of Patients With Neurologic Disease
OBJECTIVES: To describe the changes in ventilator management over time in patients with neurologic disease at ICU admission and to estimate factors associated with 28-day hospital mortality. DESIGN: Secondary analysis of three prospective, observational, multicenter studies. SETTING: Cohort studies conducted in 2004, 2010, and 2016. PATIENTS: Adult patients who received mechanical ventilation for more than 12 hours. INTERVENTIONS: None. MEASUREMENTS AND MAIN RESULTS: Among the 20,929 patients enrolled, we included 4,152 (20%) mechanically ventilated patients due to different neurologic diseases. Hemorrhagic stroke and brain trauma were the most common pathologies associated with the need fo…