19F-MRT der Lungenventilation in Atemanhaltetechnik mittels SF6-Gas
Breathhold 19 F-Magnetic Resonance Imaging of Lung Ventilation using SF 6 Gas. Objective: Development of a method to analyze lung ventilation by 19 F-magnetic resonance imaging (MRI) of inspired SF 6 gas during breathhold. Material and Methods: Measurements were performed with a Siemens Magnetom Vision 1.5T scanner using the conventional gradient overdrive. Coronal images of the lung were acquired using ultrafast gradient-echo pulse sequences with TR/TE/α = 1.4 ms/ 0.48 ms/40° without slice selection. With NEX = 200 averages and MA = 32 × 64 raw data matrix, the acquisition time was 9s/image. Higher spatial resolution of 4.7 × 6.3 × 15 mm 3 was obtained with a three-dimensional pulse sequen…