Depth of /spl gamma/-ray interaction within continuous crystals from the width of its scintillation light-distribution
We have studied a new and inexpensive method of measuring the depth of interaction (DOI) in /spl gamma/-ray detectors with large-sized scintillation crystals. This method takes advantage of the strong correlation between the width of the undisturbed light-distribution in continuous crystals and the /spl gamma/-ray's DOI. In order to quantify the dependence of the distribution's width with respect to the DOI, we first studied an analytical model of the light-distribution and tested it by means of Monte Carlo (MC) simulations of the light transport inside the crystal. Further we present an inexpensive modification of the commonly used charge division circuit that allows analog and instantaneo…