Abstract In this paper, an automatic method for the screening of water samples containing Cu(II) was proposed, based on peryoxalate chemiluminescence reaction using coproporphyrin I as fluorophor compound to provide selectivity and a simple flow injection (FI) chemiluminescence detector (CLD). FI system conditions were chosen in order to distinguish samples over or under legislation limit established (50 μg l−1) with high reliability. The detection limit found was 9 μg l−1 and the linear dynamic range was 15–125 μg l−1 of Cu(II). Repeatibility and reproducibility studies gave good precision and accuracy with recovery near 100%. Under these conditions, the method resulted selective and only …

Detection limitReproducibilityAccuracy and precisionChemistryAnalytical chemistrychemistry.chemical_elementCopperAnalytical Chemistrylaw.inventionchemistry.chemical_compoundlawTCPOSelectivityHydrogen peroxideChemiluminescenceTalanta
researchProduct