Dosimetry for electron Intra-Operative RadioTherapy: Comparison of output factors obtained through alanine/EPR pellets, ionization chamber and Monte …

2015

Abstract In this work a comparison between the response of alanine and Markus ionization chamber was carried out for measurements of the output factors (OF) of electron beams produced by a linear accelerator used for Intra-Operative Radiation Therapy (IORT). Output factors (OF) for conventional high-energy electron beams are normally measured using ionization chamber according to international dosimetry protocols. However, the electron beams used in IORT have characteristics of dose per pulse, energy spectrum and angular distribution quite different from beams usually used in external radiotherapy, so the direct application of international dosimetry protocols may introduce additional uncer…

Nuclear and High Energy PhysicsDosimeterPhysics::Instrumentation and DetectorsChemistryIORTSettore FIS/01 - Fisica SperimentalePhysics::Medical PhysicsMonte Carlo methodDetectorGeant4Dose profileElectronalanine dosimetry Monte Carlo Simulation Geant4Alanine dosimetrySettore FIS/07 - Fisica Applicata(Beni Culturali Ambientali Biol.e Medicin)Linear particle acceleratorMonte Carlo simulationsComputational physicsNuclear magnetic resonanceIonization chamberDosimetrySettore MED/36 - Diagnostica Per Immagini E RadioterapiaInstrumentationNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms