0000000000441217

AUTHOR

Rebekka Scheifl

showing 1 related works from this author

Are there any differences in various polyaxial locking systems? A mechanical study of different locking screws in multidirectional angular stable dis…

2012

Numerous angular stable plates for the distal radius exist, and technically based comparisons of the polyaxial locking interfaces are lacking. The aim of this mechanical study was to investigate three different locking interfaces of angular stable volar plates by cantilever bending: VA-LCP Two-Column Distal Radius Plates 2.4 mm (Synthes® GmbH, Oberdorf, Switzerland), IXOS® P4 (Martin, Tuttlingen, Germany) and VariAX™ (Stryker®, Duisburg, Germany). We assessed the strength of 0°, 5°, 10° and 15° screw locking angles and tested the bending strength from 10° to 5° angles by cyclic loading until breakage. The final setup repeated the above assessments by inclusion of four locking screws. The si…

Compressive StrengthFrictionbusiness.industryBone ScrewsBiomedical EngineeringRadiusStructural engineeringProsthesis DesignInternal FixatorsMechanism (engineering)Equipment Failure AnalysisCompressive strengthFlexural strengthBreakageElastic ModulusTensile StrengthBone plateUltimate tensile strengthBending momentHumansbusinessRadius FracturesBone PlatesBiomedizinische Technik. Biomedical engineering
researchProduct