PERIPHERAL SURFACE DOSE FROM A LINEAR ACCELERATOR: RADIOCHROMIC FILM EXPERIMENTAL MEASUREMENTS OF FLATTENING FILTER FREE VERSUS FLATTENED BEAMS
Abstract There is a growing interest in the use of flattening filter free (FFF) beams due to the shorter treatment times. The reduction of head scatter suggests a better radiation protection to radiotherapy patients, considering the expected decrease in peripheral surface dose (PSD). In this work, PSD of flattened (FF) and FFF-photon beams was compared. A radiochromic film calibration method to reduce energy dependence was used. PSD was measured at distances from 2 to 50 cm to the field border for different square field sizes, modifying relevant clinical parameters. Also, clinical breast and prostate stereotactic body radiotherapy (SBRT) plans were studied. For square beams, FFF PSD is lowe…