Vector polarimetry at MaMi – Measurements of tensor correlation coefficients in e−bremsstrahlungs processes
Electron/photon tensor-correlation coefficients may allow to design a polarimeter that can measure all components of beam polarization simultaneously—a so-called vector polarimeter. Besides its purpose as a beam diagnostic device this vector polarimeter would also allow to test theoretical predictions for the electron-photon polarization correlations at energies between 1 MeV and 3.5 MeV. As a first step we have set up a measurement of the helicity transfer to the photon as a function of energy which is based on the Compton absorption method.
Compton transmission polarimeter for a very precise polarization measurement within a wide range of electron currents
For new experiments with polarized electron and positron beams the precise and quick measurement of the beam polarization is required. The relative electron polarization has been monitored on-line using a compact transmission Compton polarimeter for a wide range of average electron currents up to 100/μA and electron beam energy of 3.5 MeV. The asymmetry induced by the beam polarization can be monitored precisely with ΔA/A < 0.2 %. Even small relative variations of the beam polarization during a beam time may be detected using this polarimeter, such as an increase ~ 1 % and subsequent decrease of the beam polarization.