Multicenter, Randomized, Controlled Trial of 150-J Biphasic Shocks Compared With 200- to 360-J Monophasic Shocks in the Resuscitation of Out-of-Hospi…

2000

Background —In the present study, we compared an automatic external defibrillator (AED) that delivers 150-J biphasic shocks with traditional high-energy (200- to 360-J) monophasic AEDs. Methods and Results —AEDs were prospectively randomized according to defibrillation waveform on a daily basis in 4 emergency medical services systems. Defibrillation efficacy, survival to hospital admission and discharge, return of spontaneous circulation, and neurological status at discharge (cerebral performance category) were compared. Of 338 patients with out-of-hospital cardiac arrest, 115 had a cardiac etiology, presented with ventricular fibrillation, and were shocked with an AED. The time from the e…

medicine.medical_specialtyResuscitationbusiness.industryDefibrillationmedicine.medical_treatmentReturn of spontaneous circulationmedicine.diseaseSurgerylaw.inventionRandomized controlled triallawPhysiology (medical)AnesthesiaVentricular fibrillationmedicineEmergency medical servicesCardiopulmonary resuscitationCardiology and Cardiovascular MedicinebusinessSurvival rateCirculation