Biomechanical evaluation of oversized drilling technique on primary implant stability measured by insertion torque and resonance frequency analysis

2015

Background This study evaluated the influence of implant site preparation depth on primary stability measured by insertion torque and resonance frequency analysis (RFA). Material and Methods Thirty-two implant sites were prepared in eight veal rib blocks. Sixteen sites were prepared using the conventional drilling sequence recommended by the manufacturer to a working depth of 10mm. The remaining 16 sites were prepared using an oversize drilling technique (overpreparation) to a working depth of 12mm. Bone density was determined using cone beam computerized tomography (CBCT). The implants were placed and primary stability was measured by two methods: insertion torque (Ncm), and RFA (implant s…

Materials scienceBone densityResearchDrilling030209 endocrinology & metabolismOdontología030206 dentistryImplant stability quotient:CIENCIAS MÉDICAS [UNESCO]Stability (probability)Ciencias de la salud03 medical and health sciencesResonance frequency analysis0302 clinical medicineUNESCO::CIENCIAS MÉDICASTorqueImplantTomographyGeneral DentistryImplantologyBiomedical engineering