Repeatability of the resonance frequency analysis values in implants with a new technology

2019

Background Assess the reliability (by means of reproducibility and repeatability) of the PenguinRFA system, analyse the ISQ values of different implant types and correlate the ISQ with the insertion torque during the placement of the implant. Material and Methods 120 rough surface implants were placed in bovine bone (type II and III). The implants were divided into groups, according to its design. Once the implants were in place, the exact insertion torque was registered. Then, primary stability was measured by means of the resonance frequency analysis with the PenguinRFA and the Osstell ISQ devices. In each implant two transducers of each device were used. Three measurements were obtained …

Materials scienceIntraclass correlationVibrationDental Prosthesis Retention03 medical and health sciences0302 clinical medicineAnimalsGeneral DentistryDental ImplantsReproducibilityResearchDental Implantation EndosseousResonance Frequency AnalysisReproducibility of Results030206 dentistryRepeatability:CIENCIAS MÉDICAS [UNESCO]Implant stability quotientResonance frequency analysisDental Prosthesis DesignTorqueOtorhinolaryngologyRough surfaceUNESCO::CIENCIAS MÉDICASImplant typesCattleSurgeryImplantOral SurgeryBiomedical engineeringMedicina Oral Patología Oral y Cirugia Bucal