One-year results from the DETOUR I trial of the PQ Bypass DETOUR System for percutaneous femoropopliteal bypass
The objective of this study was to evaluate the 1-year safety and effectiveness outcomes associated with the PQ Bypass DETOUR System (PQ Bypass, Milpitas, Calif) for the percutaneous bypass of long-segment femoropopliteal occlusive disease.This prospective, single-arm, multicenter trial enrolled patients with long-segment femoropopliteal arterial disease. The DETOUR System percutaneously deploys modular stent grafts to bypass femoropopliteal lesions through a transvenous route. Eligible patients included those with TransAtlantic Inter-Society Consensus C and D lesions100 mm in length. The primary safety end point was the major adverse event (MAE) rate through 1 month, defined as the composi…