Intraoperative C-arm CT imaging in angular stable plate osteosynthesis of distal radius fractures

2013

The purpose of this study was to analyze the practicability and benefit of intraoperative C-arm computed tomography (CT) imaging in volar plate osteosynthesis of unstable distal radius fractures. During a 1 year period, intraoperative three dimensional (3D) imaging with the ARCADIS Orbic 3D was performed in addition to standard fluoroscopy in 51 cases. The volar angular stable plate oesteosyntheses were analyzed intraoperatively and, if necessary, improved immediately. The duration of the scan and radiation exposure dose were measured. On average, performance of the scan and analysis of the CT dataset took 6.7 minutes. In 31.3% of the surgeries a misplacement of screws was detected and cor…

AdultMalemedicine.medical_specialtyBone ScrewsComputed tomographyRadiation DosageFracture Fixation InternalIntraoperative PeriodImaging Three-DimensionalmedicineHumansFluoroscopyProspective StudiesAgedDigital radiographyAged 80 and overOsteosynthesismedicine.diagnostic_testbusiness.industryRadiusMiddle AgedRadiation exposureTreatment OutcomePlate osteosynthesisFluoroscopyFemaleSurgeryRadiologyCt imagingRadius FracturesTomography X-Ray ComputedbusinessBone PlatesJournal of Hand Surgery (European Volume)
researchProduct