Immunonephelometric determination of the C4b-binding protein.
A fully mechanised immunonephelometric method for the rapid and specific determination of C4b-binding protein (C4b-BP) in citrated plasma is described. The method utilizes commercially available rabbit antiserum against human C4b-BP and a nephelometer analyser. A single determination can be performed within 6 min, requiring 80 microliters sample volume. The measuring range is about 10 to 200% of normal C4b-BP. Precision is characterized by intraassay coefficients of variation between 1.5% and 2.8%, and interassay coefficients of variation between 4.0% and 4.6% for the same C4b-BP concentrations. The nephelometry of C4b-BP was correlated with electroimmunodiffusion (Laurell technique; r = 0.…