Particle image velocimetry study of the celiac trunk hemodynamic induced by continuous-flow left ventricular assist device
Abstract Whereas left ventricular assist device (LVAD) is the gold-standard therapy for patients with heart failure, gastrointestinal bleeding is one of the most common complications. LVAD implantation may remarkably impact aortic hemodynamics so that experimental and computational flow analyses can be used to study the disease mechanisms. Here we present an experimentally-calibrated computational model of the celiac trunk hemodynamic of a LVAD-supported patient who experienced bleeding after device implantation. Specifically, both particle image velocimetry (PIV) and echocardiography were used to measure and compare flow distributions in each branch of a phantom model of the patient abdomi…