0000000000953835

AUTHOR

Sara Ferrer-altabás

0000-0003-0238-117x

Vectofocimetry: Dioptric power measurement in manual focimeters using power vectors

Abstract Dioptric power measurement is classically achieved using manual focimeters. In this contribution, we present vectofocimetry as a novel methodology to be easily implemented in manual focimeters for improving their capabilities regarding dioptric power measurement. The new procedure is conceptually based on power vector formalism of dioptric power which represents the dioptric power as three values in the form of [M, J0, J45], it uses an iterative process of blur reduction, and it only requires a few modifications in the device that are easy-to-adapt and cost-effective. As a result, dioptric power measurements are easily and quickly obtained with comparable results to the reported on…

research product

Characterization of a compact low-cost Stokes lens for astigmatism compensation in optical instruments

Variable power cross-cylinder lenses (or Stokes lenses) have been widely known in the literature for decades. In this paper, we describe how to build a low-cost Stokes lens and discuss its calibration and its application to two significant cases. The construction is in virtue of a phoropter’s Risley prism mount for assembling a couple of equal but opposite sign cylindrical lenses (we have selected ± 1.50 D). Thus, variable astigmatic power is achieved by relative rotation of the lenses in opposite directions, and the resulting astigmatic axis is defined by the global rotation of the device. Calibration measurements are performed using an automatic lensmeter (Topcon CL-300) and an aberromete…

research product