Fábio R.p. Rocha

A multicommuted flow system with solenoid micro-pumps for paraquat determination in natural waters.

A flow system designed with solenoid micro-pumps is proposed for the determination of paraquat in natural waters. The procedure involves the reaction of paraquat with dehydroascorbic acid followed by spectrophotometric measurements. The proposed procedure minimizes the main drawbacks related to the standard chromatographic procedure and to flow analysis and manual methods with spectrophotometric detection based on the reaction with sodium dithionite, i.e. high solvent consumption and waste generation and low sampling rate for chromatography and high instability of the reagent in the spectrophotometric procedures. A home-made 10-cm optical-path flow cell was employed for improving sensitivit…

research product

A clean method for flow injection spectrophotometric determination of cyclamate in table sweeteners

Abstract A flow system based on the multicommutation is proposed for fast and clean determination of cyclamate. The procedure exploits the reaction of cyclamate with nitrite in acidic medium and the spectrophotometric determination of the excess of nitrite by iodometry. The flow system was designed with a set of solenoid micro-pumps to minimize reagent consumption and waste generation. The detection limit was estimated as 30 μmol L −1 (99.7% confidence level) with linear response ranging up to 3.0 mmol L −1 . The coefficient of variation was estimated as 1.7% for a solution containing 2.0 mmol L −1 cyclamate ( n  = 20). About 60 samples can be analyzed per hour, consuming only 3 mg KI and 1…

research product

A portable and low cost equipment for flow injection chemiluminescence measurements.

A compact, reliable and low cost flow injection chemiluminescence system is described. The flow system consists of a set of solenoid micro-pumps that can dispense reproductive micro-volumes of solutions. The luminometer was based on a coiled cell constructed from polyethylene tubing that was sandwiched between two large area photodiodes. The whole equipment costs about US$ 750 and weights ca. 3 kg. Equipment performance was evaluated by measuring low concentrations of hydrogen peroxide by oxidation of luminol and for the determination of ammonium, based on its inhibition of the luminescence provided by the reaction of luminol and sodium hypochlorite. Linear responses were achieved within 1.…

research product