Effect of Simulated IOL Tilt and Decentration on Spherical Aberration After Hyperopic LASIK for Different Intraocular Lenses
PURPOSE: To evaluate visual quality differences among intraocular lenses (IOLs) in patients with previous hyperopic laser ablations and to assess the impact of decentration and tilt of IOLs on visual quality. METHODS: An adaptive optics visual simulator was used to simulate the wavefront aberration pattern of one aberration-correcting IOL (AcrySof IQ SN60WF, Alcon Laboratories Inc) and two spherical IOLs with different amounts of positive spherical aberration (Akreos Adapt [Bausch & Lomb] and Triplato [AJL Ophthalmic]) in five situations—centered, 0.2 mm and 0.4 mm of decentration, and 2° and 4° of tilt—in two groups: simulated low hyperopic laser corneal ablation (low hyperopia group)…
Comparison of the Ultraviolet Light Filtering across Different Intraocular Lenses.
SIGNIFICANCE We have analyzed the ultraviolet transmittance of some commercial intraocular lenses (IOLs). The results show differences of wavelength cutoff among them. PURPOSE The purpose of this study was to measure and compare the ultraviolet light transmittance of different IOLs made out of acrylic hydrophobic, hydrophilic, and hydrophilic with hydrophobic surface materials from different manufacturers. METHODS The spectral transmission curves of eight monofocal IOLs with the same dioptric power of +20.0 diopters were measured using a PerkinElmer Lambda 35 ultraviolet/visible spectrometer. Two IOLs of each type were tested three times. The ultraviolet cutoff wavelength at 10% transmissio…