0000000001088957
AUTHOR
Marjorie Funk
showing 1 related works from this author
Testing physiologic monitor alarm customization software to reduce alarm rates and improve nurses’ experience of alarms in a medical intensive care u…
2018
BACKGROUND:Clinicians in intensive care units experience alarm fatigue related to frequent false and non-actionable alarms produced by physiologic monitors. To reduce non-actionable alarms, alarm settings may need to be customized for individual patients; however, nurses may not customize alarms because of competing demands and alarm fatigue. OBJECTIVE:To examine the effectiveness and acceptance of physiologic monitor software to support customization of alarms. METHODS:This pre/post intervention study was conducted in a 56-bed medical intensive care unit. IntelliVue® Alarm Advisor customization support software for alarm limit violations was installed on all monitors and education on its u…