Performance evaluation of a high resolution dedicated breast PET scanner

2016

Purpose: Early stage breast cancers may not be visible on a whole-body PET scan. To overcome whole-body PET limitations, several dedicated breast positron emission tomography (DbPET) systems have emerged nowadays aiming to improve spatial resolution. In this work the authors evaluate the performance of a high resolution dedicated breast PET scanner (Mammi-PET, Oncovision). Methods: Global status, uniformity, sensitivity, energy, and spatial resolution were measured. Spheres of different sizes (2.5, 4, 5, and 6 mm diameter) and various 18 fluorodeoxyglucose (18F-FDG) activity concentrations were randomly inserted in a gelatine breast phantom developed at our institution. Several lesion-to-ba…

FluorodeoxyglucoseMaterials sciencemedicine.diagnostic_testbusiness.industryResolution (electron density)General MedicineIterative reconstructioncomputer.software_genre030218 nuclear medicine & medical imaging03 medical and health sciences0302 clinical medicinePositron emission tomographyVoxel030220 oncology & carcinogenesisPet scannermedicineImage sensorNuclear medicinebusinesscomputerImage resolutionmedicine.drugMedical Physics