New balloon catheter for prolonged percutaneous transluminal coronary angioplasty and bypass flow in occluded vessels.

1986

A new balloon catheter was developed for continuous perfusion of coronary arteries during angioplasty (CPC catheter). Steerable Gruntzig balloon catheters (3.7 mm) with two lumina were formed. The first lumen was used for balloon inflation. Side holes to the second lumen proximally and distally to the balloon were created for coronary perfusion even during inflation phase. At a perfusion pressure of 120 mmHg, a flow rate of 63 +/- 3 ml/min with 0.9% saline and 43 +/- 1 ml/min with plasma expander were measured. In experiments on five dogs, dilation time until appearance of signs of ischemia could be prolonged in three of five dogs from 30 to 40 s, 120 to 203 s, and 180 to 420 s comparing co…

AdultMalemedicine.medical_specialtyTime Factorsmedicine.medical_treatmentLumen (anatomy)Arterial Occlusive DiseasesCoronary DiseaseBalloonAnginaDogsAngioplastyInternal medicineMedicineAnimalsHumansbusiness.industryBalloon catheterMiddle Agedmedicine.diseaseSurgeryCoronary arteriesPerfusionCathetermedicine.anatomical_structureCoronary vesselCardiologyRegression AnalysisCardiology and Cardiovascular MedicinebusinessAngioplasty BalloonCatheterization and cardiovascular diagnosis