Extending the applicability of an open-ring trap to perform experiments with a single laser-cooled ion.
An open-ring ion trap, also referred to as transparent trap was initially built up to perform $\beta$-$\nu$ correlation experiments with radioactive ions. This trap geometry is also well suited to perform experiments with laser-cooled ions, serving for the development of a new type of Penning trap, in the framework of the project TRAPSENSOR at the University of Granada. The goal of this project is to use a single $^{40}$Ca$^+$ ion as detector for single-ion mass spectrometry. Within this project and without any modification to the initial electrode configuration, it was possible to perform Doppler cooling on $^{40}$Ca$^+$ ions, starting from large clouds and reaching single ion sensitivity.…