Comparison of two different devices to assess intraocular lenses
[EN] In this paper, we have compared the performance of two commercial systems, Kaleo-I and IOLA Plus, in the characterization of monofocal intraocular lenses (IOLs) according to the International Standard requirements (ISO 11979-2). The dioptric power (DP) and the Modulation Transfer Function (MTF) at 100.1p/mm and 3.0 mm aperture have been measured with both instruments in twenty commercial monofocal HEMA IOLs and their agreement was assessed by Bland Altman analysis. Compared with the designed power, IOLA Plus gave lower values in 85% of measures: Kaleo I errors are dependent on the power of the lens: measurements were higher than the designed power in the low-medium power range and lowe…