Determination of phenolic compounds in surface waters using high-performance liquid chromatography and molecularly imprinted solid-phase extraction
An effective and reliable analytical method based on molecularly imprinted solid-phase extraction (MI-SPE) and high-performance liquid chromatography with photodiode array detector was developed to simultaneously determine four phenolic compounds: bisphenol A, 2-phenylphenol, tert-octylphenol and nonylphenol. A molecularly imprinted polymer that can selectively recognize bisphenol A and alkylphenols was developed. A bulk polymerization method using bisphenol A as a template, acrylamide as functional monomer, ethylene glycol dimethacrylate as the cross-linking agent, acetonitrile as porogenic solvent and 1,1’-azobis(cyclohexanecarbonitrile) as initiator was applied. This polymer was applied …