A comparison of the Canon TX-20P™ non-contact tonometer and pachymeter in healthy eyes
Background: The aim of this study was to compare the intraocular pressure (IOP) and the central corneal thickness (CCT) measurements obtained with the Canon TX-20P™ with IOP measured with Goldmann applanation tonometer (GAT) and the CCT measured with ultrasound (US) pachymetry in healthy eyes. Methods: Thirty one healthy voluntary subjects participated in the study, age ranging from 16 to 81 years (mean age 50.9 years). In total 58 eyes were selected for analysis. All measurements were performed by the same examiner. The measurements were taken in the same order in all subjects: Canon IOP and CCT, Goldmann IOP and US CCT. Results: A strong correlation was found between IOP for both CCT comp…