0000000001206178

AUTHOR

Juan Agüero

0000-0002-3882-5521

showing 1 related works from this author

Validation of a high-performance chromatographic method for determination of cefotaxime in biological samples

1999

An analytical method for detecting and quantifying cefotaxime in plasma and several tissues is described. The method was developed and validated using plasma and tissues of rats. The samples were analyzed by reversed phase liquid chromatography (HPLC) with UV detection (254 nm). Calibration graphs showed a linear correlation (r > 0.999) over the concentration ranges of 0.5–200 μg/mL and 1.25–25 μg/g for plasma and tissues, respectively. The recovery of cefotaxime from plasma standards prepared at the concentrations of 25 μg/mL and 100 μg/mL was 98.5 ± 3.5% and 101.8 ± 2.2%, respectively. The recovery of cefotaxime from tissue standards of liver, fat and muscle, prepared at the concentration…

ChromatographyCefotaximePharmacokineticsChemistryCoefficient of variationBlood plasmamedicineReversed-phase chromatographyBiochemistryQuantitative analysis (chemistry)High-performance liquid chromatographyAntibacterial agentmedicine.drugFresenius' Journal of Analytical Chemistry