The ion trap aerosol mass spectrometer: field intercomparison with the ToF-AMS and the capability of differentiating organic compound classes via MS-MS
Further development and optimisation of a previously described ion trap aerosol mass spectrometer (IT-AMS) are presented, which resulted in more reproducible and robust operation and allowed for the instrument's first field deployment. Results from this 11-day-long measurement indicate that the instrument is capable of providing quantitative information on organics, nitrate, and sulfate mass concentrations with reasonable detection limits (0.5–1.4 µg m−3 for 1 h averages) and that results obtained with the IT-AMS can directly be related to those from Aerodyne aerosol mass spectrometers. The capability of the IT-AMS to elucidate the structure of fragment ions is demonstrated via an MS4 study…