Effects of minute misregistrations of prefabricated markers for image-guided dental implant surgery: an analytical evaluation

2012

Objectives The goal of the present study was to develop a theoretical analysis of errors in implant position, which can occur owing to minute registration errors of a reference marker in a cone beam computed tomography volume when inserting an implant with a surgical stent. Material and methods A virtual dental-arch model was created using anatomic data derived from the literature. Basic trigonometry was used to compute effects of defined minute registration errors of only voxel size. The errors occurring at the implant's neck and apex both in horizontal as in vertical direction were computed for mean ±95%-confidence intervals of jaw width and length and typical implant lengths (8, 10 and 1…

Dental ImplantsCone beam computed tomographymedicine.medical_specialtyComputer scienceDental Implantation EndosseousCone-Beam Computed Tomographycomputer.software_genreModels DentalPatient Care PlanningSurgeryDental implant surgeryImaging Three-DimensionalDental Prosthesis DesignSurgery Computer-AssistedVoxelPosition (vector)Vertical directionRange (statistics)medicineHumansImplantOral SurgeryFrontal regioncomputerBiomedical engineeringClinical Oral Implants Research