EPR/alanine dosimetry for verification in Helical Tomotherapy Stereotactic Radiosurgery (HTSRS) treatments

Introduction Intracranial stereotactic radiosurgery (SRS) is a technique to deliver an ablative radiation dose with an extremely sharp dose gradient to small brain tumors. In this study the accuracy of the dose delivered in SRS by a non conventional radiotherapy machine, the TomoTherapy Hi-Art System, was investigated using an "end-to-end" test using alanine pellets and gafchromic films. Methods Dose verifications were made using alanine dosimeters placed in an antropomorphic head phantom (Alderson Rando Phantom) under different treatment conditions in case of both single and multiple brain tumors. 1.25mm slice kVCT scan of the phantom was used to generate SRS plans on the TomoTherapy Plann…

research product

Characterization of phenolic solid state pellets for ESR dosimetry with radio-therapeutic photon and electron beams

Introduction Among the various dosimetric techniques used for characterizing the radiation beams used in radiation therapy, the electron spin resonance (ESR) arouses increasing interest for applications in various therapy procedures. In this work we report the ESR investigation of particular phenol compound (IRGANOX 1076) exposed to clinical photon and electron beams (Gallo et al., 2017). Methods Phenol (IRGANOX 1076 - Sigma Aldrich) pellets were produced also with paraffin (10% by weight). Phenol pellets were exposed to clinical photon and electron beams at various energies produced by a linear accelerator (LINAC) Siemens Primus (Siemens Medical Systems, CA, USA) installed at the Radiother…