Imaging of Ventriculoperitoneal Shunt Complications: Comparison of Whole Body Low-Dose Computed Tomography and Radiographic Shunt Series.

2016

OBJECTIVE To determine diagnostic value and radiation exposure of low-dose computed tomography (LD-CT) compared to radiographic shunt series (SS) for the detection of ventriculoperitoneal (VP) shunt complications. METHODS Fourteen VP shunts were implanted in 7 swine cadavers. Mechanical complications were induced in 50% of VP shunts. Low-dose CT (80 kVp, 10 mAs, Pitch = 1.5) and SS were acquired. Dose area product (DAP) and effective doses for SS and LD-CT were collected. Scoring of diagnostic confidence and blinded readings of SS and CT data were performed. RESULTS The sensitivity of LD-CT was high (0.97; 95% confidence interval, 0.91-1.00) with excellent interobserver agreement (κ = 0.88)…

medicine.medical_specialtySwineRadiographyComputed tomographyRadiation DosageEffective dose (radiation)Sensitivity and SpecificityVentriculoperitoneal Shunt030218 nuclear medicine & medical imaging03 medical and health sciences0302 clinical medicinePostoperative ComplicationsRadiation ProtectionmedicineAnimalsRadiology Nuclear Medicine and imagingMedical physicsSingle-Blind MethodWhole Body Imagingmedicine.diagnostic_testbusiness.industryLow doseReproducibility of ResultsRadiation ExposureConfidence intervalRadiographic Image EnhancementDose area productRadiographic Image Interpretation Computer-AssistedTomographybusinessNuclear medicineTomography X-Ray Computed030217 neurology & neurosurgeryShunt (electrical)Journal of computer assisted tomography