The migration of femoral components after total hip replacement surgery: accuracy and precision of software-aided measurements.
To assess the accuracy and precision of a software-aided system to measure migration of femoral components after total hip replacement (THR) on digitised radiographs.Subsidence and varus-valgus tilt of THR stems within the femur were measured on digitised anteroposterior pelvic radiographs. The measuring software (UMA, GEMED, Germany) relies on bony landmarks and comparability parameters of two consecutive radiographs. Its accuracy and precision were calculated by comparing it with the gold standard in migration measurements, radiostereometric analysis (RSA). Radiographs and corresponding RSA measurements were performed in 60 patients (38-69 years) following cementless THR surgery.The UMA s…