Application of an O-ring pinch device as a constant pressure inlet (CPI) for airborne sampling
Abstract. We present a novel and compact design of a constant pressure inlet (CPI) developed for use in airborne aerosol mass spectrometry. In particular the inlet system is optimized for aerodynamic lenses commonly used in aerosol mass spectrometers, where efficient focusing of aerosol particles into a vacuum chamber requires a precisely controlled lens pressure, typically of a few hPa. The CPI device can also be used in gas phase sampling instruments in a large range of altitude and inlet pressure. The constant pressure is achieved by changing the inner diameter of a properly scaled O-ring that acts as a critical orifice. The CPI control keeps air pressure and thereby mass flow rate (≈ 0.…
Application of an O-ring pinch device as a constant-pressure inlet (CPI) for airborne sampling
We present a novel and compact design of a constant-pressure inlet (CPI) developed for use in airborne aerosol mass spectrometry. In particular, the inlet system is optimized for aerodynamic lenses commonly used in aerosol mass spectrometers, in which efficient focusing of aerosol particles into a vacuum chamber requires a precisely controlled lens pressure, typically of a few hectopascals. The CPI device can also be used in condensation particle counters (CPCs), cloud condensation nucleus counters (CCNCs), and gas-phase sampling instruments across a wide range of altitudes and inlet pressures. The constant pressure is achieved by changing the inner diameter of a properly scaled O-ring that…
Application of an O-ring pinch device as a constant pressure inlet (CPI) for airborne sampling
We present a novel and compact design of a constant pressure inlet (CPI) developed for use in airborne aerosol mass spectrometry. In particular the inlet system is optimized for aerodynamic lenses commonly used in aerosol mass spectrometers, where efficient focusing of aerosol particles into a vacuum chamber requires a precisely controlled lens pressure, typically of a few hPa. The CPI device can also be used in gas phase sampling instruments in a large range of altitude and inlet pressure. The constant pressure is achieved by changing the inner diameter of a properly scaled O-ring that acts as a critical orifice. The CPI control keeps air pressure and thereby mass flow rate (≈&thinsp…