6533b822fe1ef96bd127ce06
RESEARCH PRODUCT
Virtual point detector: Application to coincidence-summing correction in gamma-ray spectrometry
Elio Angelo TomarchioSalvatore Rizzosubject
PhysicsRadiationbusiness.industrySettore ING-IND/20 - Misure E Strumentazione NucleariPoint detectorGamma-ray spectrometryCoincidence-summingCoincidenceOpticsVolume (thermodynamics)Point (geometry)Virtual point detectorbusinessGamma ray spectrometrydescription
The use of an efficiency transfer method based on the virtual point detector approximation to evaluate coincidence-summing correction values for point and volume sources was examined. The method was validated by means of a comparison with experimental efficiency determinations.
year | journal | country | edition | language |
---|---|---|---|---|
2009-12-01 |