6533b825fe1ef96bd128278b
RESEARCH PRODUCT
Proximale Verankerung von Hüft-Endoprothesen mit poröser Oberfläche: 6-Jahres-Ergebnisse
D. DeckingJ. DeckingT. Rosendahlsubject
medicine.medical_specialtyPorous coatingChirurgie orthopediquebusiness.industryRadiographyTotal hip replacementequipment and suppliesSurgeryOrthopedic surgeryThigh painMedicineOrthopedics and Sports MedicineSurgeryFemurbusinessNuclear medicineFixation (histology)description
PROBLEM The anatomically shaped pcl (ESKA) total hip replacement has a porous coating of the proximal stem and the cup to allow for bony ingrowth. The tip of the stem is polished. METHOD We report the results of 96 THR 6.2 years (5.25-6.7) after implantation. RESULTS There were no revisions during this period. One stem showed radiographic loosening at follow-up. There was stress-shielding in 9.4% in contrast to the concept of proximal fixation. Thigh-pain was found in 9.4%. One patient demonstrated disabling thigh-pain despite radiologic bony ingrowth. By drilling the femur distal to the tip of the stem a remarkable relief was achieved. CONCLUSION Despite numerous cases of stress-shielding (9.4%), bony ingrowth occurs in 97%. The tcl-THR performs sufficiently after 6 years of implantation.
year | journal | country | edition | language |
---|---|---|---|---|
2008-05-14 | Zeitschrift für Orthopädie und ihre Grenzgebiete |