FIA-fluorimetric determination of adrenaline in pharmaceutical formulations by oxidation with molecular oxygen
A. Mellado RomeroA. Canoves TorresJ. Martínez Calatayudsubject
Aqueous solutionChromatographyReducing agentChemistryFluorometerFluorescence spectrometryMolecular oxygenQuantitative analysis (chemistry)Dosage formFluorescence spectroscopyAnalytical Chemistrydescription
The fluorimetric determination of adrenaline is carried out in a continuous-flow assembly and by means of the molecular dissolved oxygen. The sample solution merges with an NaOH stream, then the resulting mixture is heated at 73 °C and led to the flow-cell of the fluorimeter. The flow-assembly is very simple and the procedure is quick (107 samples h−1) reproducible (R.S.D. 0.6%), selective and suitable to be applied to determination of adrenaline in formulations. Calibrations graph are linear over the ranges 0.05–15 and 20–40 mgl−1.