6533b827fe1ef96bd1285be1
RESEARCH PRODUCT
Study of three urethral pressure recording devices: Theoretical considerations
Udo JonasHans J. Klottersubject
Malemedicine.medical_specialtyManometrybusiness.industryUrologyUrinary BladderBalloon catheterBalloonBiomechanical PhenomenaSurgeryModels StructuralCatheterUrethraUrethral pressureHydrostatic PressurePressuremedicineHumansUrinary CatheterizationbusinessTip catheterBiomedical engineeringdescription
Using a bladder-urethra model, 2 perfusion catheters (open side and open tip) were compared to the balloon catheter. The most accurate results were obtained using the balloon catheter. Based on the law of continuity and Bernouilli's law, the disadvantage of the open side catheter is due to the inconstant diameter of the sum of the side holes. Therefore, the measured pressure is not necessarily identical to the real pressure. The open tip catheter measures the opening pressure and the recorded pressure does not necessarily reflect the pressure at the site of the tip hole. The balloon catheter has elastic and plastic characteristics. Optimum results were obtained only with prestretch of the balloon and with calibration before and after each study. This catheter was the most complicated, but produced the best results and gave rise to the least irritation.
year | journal | country | edition | language |
---|---|---|---|---|
1978-01-01 | Urological Research |