For high-resolution alpha-particle spectroscopy, detector calibration sources of highest performance are required concerning thickness and homogeneity. In addition, the alpha-emitting radionuclides must have adequate half-lives and alpha-emission energies. This paper describes a new type of alpha-emitting calibration source, represented by a mixture of 148Gd and 244Cm. This system covers an alpha-energy range from 3.18 MeV (148Gd) to 5.76 and 5.80 MeV (244Cm). The source has been prepared by electrolytic deposition and has good performance in terms of thickness and homogeneity.
year | journal | country | edition | language |
---|---|---|---|---|
2005-04-07 | Applied Radiation and Isotopes |