6533b834fe1ef96bd129d8bd

RESEARCH PRODUCT

A comparative study of flame atomic-absorption methods for determination of zinc in serum and blood plasma

M. De La GuardiaM J Soriano

subject

Detection limitChemistrylawBlood plasmaAnalytical chemistrychemistry.chemical_elementZincAtomic absorption spectroscopySpectroscopyBiological fluidAnalytical Chemistrylaw.invention

description

Seven selected methods for determination of zinc in blood plasma by flame atomic-absorption spectroscopy have been compared. Analytical characteristics such as sensitivity, detection limit, precision, analytical recovery, accuracy and physical interferences were studied. Two of the seven methods are recommended as the most suitable for the purpose.

https://doi.org/10.1016/0039-9140(84)80094-6