6533b85dfe1ef96bd12bdfe5
RESEARCH PRODUCT
Example of a technique for evaluation of interferences caused by complicated sample matrix elements in ICP-AES determination
Päivi HautaniemiRose Matilainensubject
Correction methodChemistryInductively coupled plasma atomic emission spectroscopyLinear regressionAnalytical chemistryMatrix elementInductively coupled plasmaError detection and correctionInterference (wave propagation)Biological systemBiochemistrySample (graphics)description
An example of a useful and rapid procedure for the evaluation of interferences caused by complicated sample matrices in inductively coupled plasma atomic emission spectrometry (ICP-AES) is described. Using simple acid-base standards, all the elements investigated were determined separately in complicated matrices with satisfactory results. Multiple linear regression was used to calculate the linear correction coefficients for each matrix element analyzed. Good analytical results improved still further when this correction method was used.
year | journal | country | edition | language |
---|---|---|---|---|
2001-05-09 | Fresenius' Journal of Analytical Chemistry |