New applicator improves waterjet dissection quality

2010

Waterjet dissection is accomplished with Helix Hydro-Jet, but a new device with improved operative handling and potentially superior dissection qualities has been developed.Eighty-four fresh cadaveric pig brains were simultaneously cut with Helix Hydro-Jet and Erbejet 2. A commonly used applicator and a new applicator for the Helix Hydro-Jet were directly compared to the new Erbejet 2. Under standardised conditions, different pressure levels were applied to the brain surface without arachnoids. Technical features, cutting depth, tissue damage and differences of applicators were examined.Microscopic analysis of cutting depth revealed different dissection characteristics of both the devices. …

medicine.medical_specialtySwineHydrostatic pressureBrain surfaceDissection (medical)Neurosurgical ProceduresTissue damageHydrostatic PressuremedicineAnimalsNew devicebusiness.industryDissectionBrainEquipment DesignGeneral MedicineSurgical Instrumentsmedicine.diseaseSurgeryArachnoidsModels AnimalSurgeryNeurology (clinical)businessCadaveric spasmBiomedical engineeringBritish Journal of Neurosurgery