Technical note: a comparison of central and peripheral intraocular pressure using rebound tonometry

2007

Purpose: To compare central and peripheral intraocular pressure (IOP) readings obtained with rebound tonometry.Methods: Intraocular pressure was measured on the right eye of 153 patients (65 males, 88 females), aged from 21 to 85 years (mean +/- S.D., 55.5 +/- 15.2 years) with the ICare rebound tonometer at centre, and 2 mm from the limbus (in the nasal and temporal regions along the 0-180O corneal meridian).Results: Intraocular pressure values obtained with the ICare were 14.9 +/- 2.8; 14.1 +/- 2.5 and 14.5 +/- 2.7 mmHg at centre, nasal and temporal corneal locations, respectively. On average, nasal and temporal IOP readings were 0.75 and 0.37 mmHg lower than the central reading (p 0.05, r…

AdultMalemedicine.medical_specialtyIntraocular pressuregenetic structuresSensitivity and SpecificityICare((R))Cornea03 medical and health sciencesTonometry Ocular0302 clinical medicineReference ValuesOphthalmologyCorneamedicineHumansIntraocular PressureAgedAged 80 and overScience & Technologybusiness.industryReproducibility of ResultsTechnical noteREBOUND TONOMETRYMiddle AgedSensory Systemseye diseasesPeripheralOphthalmologymedicine.anatomical_structureMeridian (perimetry visual field)Temporal RegionsReference values030221 ophthalmology & optometryFemalerebound tonometrysense organsICare®corneal structureperipheral tonometrybusiness030217 neurology & neurosurgeryOptometryintraocular pressure