Refractive analysis of the human cornea through propagated fields

2001

A new technique for analysing the optical quality of the human cornea is presented. Corneal maps are obtained through keratographies and then converted into phase maps. The propagated fields generated from this surface are plotted and studied. It is shown that any irregularity in the cornea affects the propagated field and the energy distribution at the focal plane. Simple applications are also indicated.

Surface (mathematics)Materials scienceEnergy distributionField (physics)Wave propagationbusiness.industryPhysics::Medical PhysicsPhase (waves)eye diseasesAtomic and Molecular Physics and OpticsOptical qualityCardinal pointmedicine.anatomical_structureOpticsCorneamedicinesense organsbusinessJournal of Modern Optics
researchProduct

In vitro optical performance of a new aberration-free intraocular lens

2014

To assess the optical performance of the new EnVista intraocular lens (IOL). Four aspheric IOLs were evaluated; the new EnVista is one amoung them. This IOL, similarly to the Z-Flex HB and the Bi-Flex 1.8 667AB ones, has a neutral aspheric design, whereas the fourth IOL under test (AcrySof IQ IOL SN60WF) presents a negative spherical aberration (SA). The IOL’s aberration patterns were measured in vitro, by setting them up on an optical bench. From these aberration-pattern data, the modulation transfer function (MTF), the average modulation values, and the points spread function (PSF) were calculated. Furthermore, in order to assess the potential optical quality that these IOLs would yield o…

WavefrontLenses Intraocularmedicine.medical_specialtyOptics and PhotonicsMaterials scienceCorneal Wavefront Aberrationgenetic structuresmedicine.medical_treatmentIntraocular lensequipment and suppliesProsthesis Designeye diseasesOptical qualityOphthalmologySpherical aberrationmedicine.anatomical_structureCorneaOphthalmologyOptical transfer functionLaboratory StudymedicineProsthesis designHumanssense organsCorneal Wavefront Aberration
researchProduct

The tear film and the optical quality of the eye

2010

The stability of the tear film plays an important role in the optical quality of the eye. Various methods have been developed to quantify the contribution of the tear film to optical quality, including the double-pass optical method, Hartmann-Shack wavefront sensing, videokeratoscopy, retroillumination analysis, and interferometry. In addition to assessing the effect of tear film irregularities, these methods can aid in the diagnosis of dry-eye related conditions and the monitoring of therapy. This review describes the methodologies used to assess the relationship between the tear film and optical quality of the eye and summarizes the most important findings obtained with these methods with…

Wavefrontmedicine.medical_specialtyCorneal Wavefront AberrationMaterials sciencegenetic structuresmedicine.medical_treatmenteye diseasesOptical qualityCorneaOphthalmologyInterferometryArtificial tearsmedicine.anatomical_structureTearsOphthalmologymedicineHumansOptometryDry Eye SyndromesHuman eyesense organs
researchProduct

An improved molecular design of obtaining NLO active molecular glasses using triphenyl moieties as amorphous phase formation enhancers

2012

New molecular design of obtaining molecular glasses has been developed by linking triphenylmethyl moieties to chromophore core by flexible C-C bridge. Compounds capable of forming stable amorphous phase with good optical quality have been acquired with increased chemical and thermal sustainability compared to the previously reported design. NLO activity of compounds has been measured after corona discharge polling. Compared to previously synthesized trityloxy fragment containing compounds increase of d 33 coefficient by up to 17 times was achieved for the same chromophore core containing compounds.

chemistry.chemical_compoundMaterials scienceAzobenzenechemistryMolecular glassesSecond-harmonic generationNonlinear opticsChromophorePhotochemistryAmorphous phaseCorona dischargeOptical qualitySPIE Proceedings
researchProduct

Modulation transfer function in refractive-diffractive multifocal intraocular lenses in vitro

2007

Purpose: The optical quality of two types of multifocal, hybrid refractive-diffractive, intraocular lenses (IOL) is analyzed in vitro. Following the EN-ISO international standard, the IOLs’ modulation transfer functions (MTF) are determined for both distance and near vision. The behaviour of these curves when the pupil size varies is studied. Methods: The IOLs analyzed were: AcrySoft ResTor SN60D3 (Alcon) and Tecnis ZM900 (AMO). The MTF was calculated from the cross line-spread function recorded with the OPAL Vector System (Image Science Ltd. Oxford) by using fast Fourier transform techniques. The artificial eye model used simulates in vivo conditions of the anterior chamber including an ar…

genetic structuresbusiness.industryImage qualityStrehl ratioMultifocal intraocular lenseye diseasesPupilOptical qualityOphthalmologyOpticsIntraocular lensesOptical transfer functionMedicinesense organsImaging sciencebusinessActa Ophthalmologica Scandinavica
researchProduct

Optical quality of intraocular lenses after several years in the human eye

2013

The optical quality of explanted intraocular lenses (IOLs) was analysed to verify if their quality is maintained after having been in the human eye for several years. Twenty-five explanted IOLs were studied, 10 because of intraocular luxation and 15 from corneal donors. The dioptric power and modulation transfer function (MTF) were determined to discover the image quality of the IOLs. The spectral transmission was also established to ascertain whether the filter against ultraviolet (UV) radiation was still effective. The results show that the MTF of implanted IOLs is maintained and therefore so is the image quality they provide. Only decay in the MTF of some of the explanted IOLs was observ…

medicine.medical_specialtyMaterials sciencegenetic structuresImage qualitybusiness.industrySpectral transmissionequipment and supplieseye diseasesAtomic and Molecular Physics and OpticsOptical qualityOpticsmedicine.anatomical_structureIntraocular lensesOphthalmologymedicineHuman eyesense organsbusinessJournal of Modern Optics
researchProduct

Influence of the tear film on optical quality of the eye

1997

The optical quality of the tear film of the eye was tested by measuring the retinal image before and after the break-up time (BUT). An objective method was used based on the evaluation of the retinal image of a point test projected onto the fovea. The loss of an entire tear film would result in a decrease in the optical quality because of corneal irregularities and the formation of an irregular tear lens after the BUT. Our results confirm the expected loss both of non-contact lens wearers and contact lens wearers. Also, the fact that the optical deterioration found after the BUT is greater for contact lens wearers confirms that soft contact lens wear produces a disruption of the tears.

medicine.medical_specialtyMaterials sciencegenetic structuresbusiness.industryObjective methodGeneral Medicineeye diseasesOptical qualityRetinal imagelaw.inventionLens (optics)Contact lensOphthalmologyOpticslawOphthalmologyMerit functionmedicineTearssense organsbusinessOptometry
researchProduct

Role of the tear film in the optical quality of the human eye.

2007

This paper reviews recent studies of the role of the tear film in the optical quality of the eye. Recent investigations focus on the dynamics of the tear film after a blink and its important role in the optical quality of the eye. The results in this review were used in an attempt to understand the optical and visual impact of tear-film breakup in normal eyes and dry eyes. It was noted that patients with dry-eye syndrome have large optical aberrations compared with those in normal eyes, which may be the cause of blurry vision associated with the syndrome. In addition, instillation of artificial tears reduces optical aberrations and thus improves the optical quality of vision in these patien…

medicine.medical_specialtygenetic structuresmedicine.medical_treatmentCorneaOcular physiologyOphthalmologymedicineHumansOcular Physiological PhenomenaOcular Physiological PhenomenaVision Ocularmedicine.diagnostic_testBlinkingbusiness.industryDry eyesCorneal TopographyCorneal topographymedicine.diseaseRefractive Errorseye diseasesSensory SystemsOptical qualityOphthalmologyArtificial tearsmedicine.anatomical_structureBlurry visionTearsSurgeryHuman eyeDry Eye Syndromessense organsOphthalmic SolutionsbusinessJournal of cataract and refractive surgery
researchProduct

New intraocular lens for achromatizing the human eye.

2007

To describe the design of a new intraocular lens (IOL) capable of correcting spherical and chromatic aberrations when implanted in the human eye.University of Murcia, Murcia, and University of Valencia, Valencia, Spain.A hybrid singlet achromatic IOL was designed. The IOL has a combination of a refractive and a diffractive surface, with 1 of the surfaces being aspherical. Optical simulations were used to model the polychromatic modulation transfer function (MTF) in pseudophakic eyes to explain the differences in optical quality afforded by the achromatic IOL. Parameters such as focus shift, optical path difference, through-focus, and robustness to tilt and decentering of achromatic IOLs wer…

medicine.medical_treatmentIntraocular lensEyeProsthesis DesignModels Biologicallaw.inventionOpticslawChromatic aberrationmedicineHumansDioptrePhysicsLenses Intraocularbusiness.industryModels TheoreticalCentrationSensory SystemsOptical qualityOphthalmologymedicine.anatomical_structureTilt (optics)Achromatic lensOptometrySurgeryHuman eyebusinessColor PerceptionJournal of cataract and refractive surgery
researchProduct

Optical quality of hyperopic and myopic phakic intraocular lenses

2013

Aims: To assess and compare the optical quality of the myopic and hyperopic implantable collamer lens (ICL) from its wavefront aberrations for different powers and pupil diameters. Settings and Design: Prospective study. Material and Methods: The wavefront aberrations of two myopic (−3 and −6 diopters (D)) and two hyperopic V4b ICLs (+3 and +6D) were measured in vitro . To assess and compare the optical quality of different powers of ICLs, we analyzed the root mean square (RMS) of total higher order aberrations (HOAs), trefoil, coma, tetrafoil, secondary astigmatism, and spherical aberration at 3- and 4.5-mm pupil. In addition, the point spread functions (PSFs) of each ICL evaluated were ca…

steep corneal curvatureintravitreallacrimal fossaRefractive errorvisual acuitygenetic structurespterygium surgerystandard automated perimetryComa (optics)BlindnessPhakic intraocular lenslaw.inventionsubconjunctival bevacizumab injectionCorneaLens Implantation Intraocularlcsh:Ophthalmologylawmodified deep anterior lamellar keratoplastyMitomycin CMaterials TestingAutogenous tissue graftingProspective StudiesDioptreoptical qualityshort wave automated perimetryexcimer lasernon-proliferative diabetic retinopathyMean deviationin vitroheavy silicone oilBevacizumabLens (optics)Aberrations of the eyeSpherical aberrationHyperopiaAntimicrobial effectOriginal ArticleTriamcinolone acetonidehuman trabecular meshwork cellsPhakic Intraocular Lensesmedicine.medical_specialtyImaging in glaucomadisability evaluationconventional silicone oiltriamcinoloneProsthesis DesignKeratoconusphotorefractive keratectomyOphthalmologygrowth factorsmedicineHumansmyopiamyopic implantable collamer lensshort fluctuationsWavefrontmacular edemaoptical coherence tomographyendophthalmitis agentsbusiness.industrybranch retinal vein occlusionretinal nerve fiber layerpterygium recurrencemedicine.diseaseeye diseasesOphthalmologypediatricwavefront aberrationsquality of lifelcsh:RE1-994bypass surgeryHyperopic implantable collamer lenslacrimal drainage systemsense organsbusinessIndian Journal of Ophthalmology
researchProduct