Digital CZT detector system for high flux energy-resolved X-ray imaging

2017

Photon counting arrays with energy resolving capabilities are recently desired for the next-generation X-ray imaging systems. In this work, we present the performance of a 2 mm thick CZT pixel detector, with pixel pitches of 500 mu m and 250 mu m, coupled to a fast and low noise ASIC (PIXIE ASIC), characterized by only the preamplifier stage. A 16-channel digital readout electronics was used to continuously digitize and process each output channel from the PIXIE ASIC, performing multi-parameter analysis (event arrival time, pulse shape, pulse height) at low and high input counting rates (ICRs). The spectroscopic response of the system to monochromatic X-ray and gamma ray sources, at both lo…

Radiology Nuclear Medicine and ImagingPhysics::Instrumentation and DetectorsPreamplifier01 natural sciences030218 nuclear medicine & medical imagingCharge sharing03 medical and health sciences0302 clinical medicineOptics0103 physical sciencesInstrumentationNuclear and High Energy PhysicPhysicsPixel010308 nuclear & particles physicsbusiness.industryDetectorSettore FIS/01 - Fisica SperimentaleGamma raySMALL PIXEL; CDTE; PERFORMANCE; RESOLUTION; PROGRESS; CTPhoton countingSettore FIS/07 - Fisica Applicata(Beni Culturali Ambientali Biol.e Medicin)Electronic Optical and Magnetic MaterialsPixieMonochromatic colorbusiness
researchProduct