Lung Function Monitoring; A Randomized Agreement Study

2016

Objective: To determine the agreement between devices and repeatability within devices of the forced expiratory volume in 1 s (FEV1), forced vital capacity (FVC), peak expiratory flow (PEF) and forced expiratory flow at 50% of FVC (FEF50) values measured using the four spirometers included in the study. Methods: 50 (24 women) participants (20-64 years of age) completed maximum forced expiratory flow manoeuvres and measurements were performed using the following devices: MasterScreen, SensorMedics, Oxycon Pro and SpiroUSB. The order of the instruments tested was randomized and blinded for both the participants and the technicians. Re-testing was conducted on a following day within 72 hours …

Pulmonary and Respiratory Medicinemedicine.medical_specialtyVital capacitySpirometersTestingArticleMean difference03 medical and health sciencesFEV1/FVC ratio0302 clinical medicineInternal medicinemedicineAdultsRepeatability030212 general & internal medicineLung functionReproducibilitybusiness.industryMeasurementsRepeatabilityrespiratory systemReproducibilityConfidence intervalrespiratory tract diseases030228 respiratory systemPhysical therapyCardiologybusinessThe Open Respiratory Medicine Journal