Failure mode and effects analysis of skin electronic brachytherapy using Esteya® unit

2016

Purpose: Esteya® (Nucletron, an Elekta company, Elekta AB, Stockholm, Sweden) is an electronic brachytherapy device used for skin cancer lesion treatment. In order to establish an adequate level of quality of treatment, a risk analysis of the Esteya treatment process has been done, following the methodology proposed by the TG-100 guidelines of the American Association of Physicists in Medicine (AAPM). Material and methods: A multidisciplinary team familiar with the treatment process was formed. This team developed a process map (PM) outlining the stages, through which a patient passed when subjected to the Esteya treatment. They identified potential failure modes (FM) and each individual FM…

medicine.medical_specialtyQuality managementFrequency of occurrenceelectronic brachytherapymedia_common.quotation_subjectmedicine.medical_treatmentBrachytherapylcsh:MedicineTG-10003 medical and health sciences0302 clinical medicineRisk analysis (business)MedicineRadiology Nuclear Medicine and imagingQuality (business)Medical physicsQAFMEAmedia_commonOriginal Paperskin cancerBrachytherapy devicebusiness.industry030503 health policy & serviceslcsh:RTreatment processEsteyaOncology030220 oncology & carcinogenesis0305 other medical sciencebusinessFailure mode and effects analysisJournal of Contemporary Brachytherapy