Compression forces of haptics of selected posterior chamber lenses
Abstract Purpose: To compare the compressive forces of the haptics of different intraocular lens (IOL) models and analyze the observed differences. Setting: Central Hospital of Central Finland and University of Jyvaskyla, Jyvaskyla, Finland. Methods: The haptics of 28 IOL models were compressed to a diameter of 9.0 mm. The compression forces were measured at 0.5 mm intervals. The conclusions were verified by numerical simulations of mechanical models of the lenses. Results: The measured forces varied between 100 and 601 mg at a diameter of 11.0 mm, 206 and 1057 mg at a diameter of 10.0 mm, and 315 and 2094 mg at a diameter of 9.0 mm. The slopes of the force curves of the three-piece lenses …