Closed-Loop Automatic Oxygen Control (CLAC) in Preterm Infants: A Randomized Controlled Trial

BACKGROUND AND OBJECTIVE: In preterm infants receiving supplemental oxygen, routine manual control (RMC) of the fraction of inspired oxygen (FIO2) is often difficult and time consuming. We developed a system for closed-loop automatic control (CLAC) of the FIO2 and demonstrated its short-term safety and efficacy in a single-center study. The objective of this study was to test the hypothesis that this system is more effective than RMC alone in maintaining arterial oxygen saturation within target levels when evaluated over 24 hours under routine conditions and with different target levels. METHODS: We performed a multicenter, randomized controlled, crossover clinical trial in 34 preterm infa…

research product

Is faster better? A randomised crossover study comparing algorithms for closed-loop automatic oxygen control

ObjectiveClosed-loop automatic control (CLAC) of the fractional inspired oxygen (FiO2) improved oxygen administration to preterm infants on respiratory support. We investigated whether a revised CLAC algorithm (CLACfast, ≤2 FiO2 adjustments/min), compared with routine manual control (RMConly), increased the proportion of time with arterial haemoglobin oxygen saturation measured by pulse oximetry within prespecified target ranges (Target%) while not being inferior to the original algorithm (CLACslow: ≤0.3 FiO2 adjustments/min).DesignUnblinded randomised controlled crossover study comparing three modes of FiO2 control in random order for 8 hours each: RMC supported by CLACfast was compared wi…